CVP catheter electrocardiography: an alternative to radiographic control after cannulation of central veins?
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1763598.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We evaluated 160 electrocardiograms taken after placement of central venous catheters (CVC) to determine their locations. Usable recordings were obtained in 154 patients. Subsequent radiographs revealed 30 misplaced catheters. Twenty-five of those were detected by CVC electrocardiograms. There were five false positive and five false negative traces. The sensitivity of CVC electrocardiography was 96%, the specificity 83.3%, and the total predictive power 93.5%. Electrocardiograms obtained from guide-wires were of significantly better quality than those from 0.9% NaCl filled catheters. The technique is accurate, safe and easy to learn. It may reduce the need for routine radiographic control to less than 10% of patients.
Medical subject headings
- Catheterization, Central Venous
- Electrocardiography