Ultrasound-guided breast biopsy using a 10-gauge self-contained vacuum-assisted device.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 17639409.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to evaluate the diagnostic value of a self-contained battery-driven vacuum-assisted breast biopsy (VABB) system for the sampling of breast masses under ultrasound guidance. Sixty-five patients with 70 lesions underwent percutaneous 10-gauge ultrasound-guided VABB using the coaxial technique. In 38 lesions, subsequent surgery and comparison of histology was performed. The remaining 32 cases were followed-up and defined as true negative after a cancer-free interval of 24 months. VABB revealed malignant histology in 28 (40%) cases. Twenty-four malignancies were confirmed after surgery. Four invasive cancers verified in VABB were not found during surgery because they were completely removed, as proved by a disease-free interval of 24 months. One cancer missed in ultrasound-guided VABB due to its small size was successfully diagnosed with stereotactic VABB and thus turned out to be false negative, resulting in an overall sensitivity of 96.6%. Forty-one patients were free of cancer during the follow-up period of at least 24 months. In conclusion, the self-contained VABB device is well suited for ultrasound-guided breast biopsies.
Medical subject headings
- Biopsy, Needle
- Breast Neoplasms
- Ultrasonography, Interventional
- Ultrasonography, Mammary