Spinal cord monitoring in patients with nonidiopathic spinal deformities using somatosensory evoked potentials.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1771465.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Seventy-nine somatosensory evoked potentials were intraoperatively recorded in 52 patients undergoing spinal surgery for nonidiopathic spinal deformities. There were 37 true-negative, 28 true-positive (a significant change in the somatosensory evoked potential related to the surgical process), and 14 false-positive (a significant change in the somatosensory evoked potential not related to a surgical event) readings. There were, however, no postoperative neurologic deficits with any of the true-positive readings and no false negatives. Spinal and subcortical somatosensory evoked potentials gave few false-positive readings. True-positive somatosensory evoked potentials occurred in 44% of the patients with neuromuscular deformities, 17% with congenital deformities, 45% with Luque instrumentation, 22% with Harrington instrumentation, and none with fusion in situ. Fifty percent of the true positives occurred while the sublaminar wires were tightened. The predictive accuracy of intraoperative spinal cord monitoring in this patient population is not high, but the sensitivity to potentially harmful surgical events is high.
Medical subject headings
- Evoked Potentials, Somatosensory
- Monitoring, Intraoperative
- Spinal Diseases
- Spinal Fusion