Ranges of movement at the shoulder complex using an electromagnetic movement sensor.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1772301.
- Also identified by PMC identifier 1004569.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
There are few published data on the ranges of available motion at the shoulder and no clinical technique for the assessment of three dimensional movement. An electromagnetic movement sensor (Isotrak), which makes three dimensional measurements of shoulder movement and measures the ranges of flexion, extension, abduction in the neutral position, abduction preceded by external rotation, and internal rotation achieved by placing the hand as far up the back as possible, was used to study 16 normal subjects. Intraobserver repeatability was good, and a study of five subjects by two observers showed no significant difference between single measurements made by each. The measured ranges were in broad agreement with published data, though they tended to be smaller. This was attributed to either the bulk of the mounting for the movement sensor or to the immobilisation of the elbow and hence the biceps muscle. Women had greater ranges, except for externally rotated abduction.
Medical subject headings
- Orthopedic Equipment
- Range of Motion, Articular
- Shoulder Joint