Continuous cardiac output measurement: arterial pressure analysis versus thermodilution technique during cardiac surgery with cardiopulmonary bypass.

Lorsomradee, S; Lorsomradee, S R; Cromheecke, S; De Hert, S G · Anaesthesia · 2007

prospective_cohort · Level II

Where this comes from

Abstract

This study compared cardiac output measured with an arterial pressure-based cardiac output measurement system and a thermodilution cardiac output measurement system. We studied 36 patients undergoing cardiac surgery with cardiopulmonary bypass. Simultaneous arterial pressure-based and thermodilution cardiac output measurements were compared before and after cardiopulmonary bypass, and after phenylephrine administration. Bland-Altman analysis showed good overall agreement between the two methods. Bias (limits of agreement) before and after cardiopulmonary bypass were - 0.21 (- 2.97-2.55) lxmin(-1) and 0.01 (- 3.79-3.81) lxmin(-1), respectively. Phenylephrine administration decreased thermodilution cardiac output by a mean (SD) of 11 (16)% and increased arterial pressure-based cardiac output by 55 (34)%. We conclude that arterial pressure-based cardiac output and thermodilution cardiac output measurement systems yield comparable results during cardiac surgery with cardiopulmonary bypass. However, after phenylephrine administration, the two measurement systems provided opposing results.