Continuous glucose monitoring in children with type 1 diabetes.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 17889075.
- Also identified by PMC identifier 2045068.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To examine the feasibility of daily use of a continuous glucose monitor, the FreeStyle Navigator Continuous Glucose Monitoring System ("Navigator"), in children with type 1 diabetes (T1D). After a masked Navigator was used for 4 to 7 days to establish a baseline level of glycemic control, 30 insulin pump users with T1D (average age 11.2 years) were asked to use the Navigator daily for 13 weeks. Subjects averaged 149 h/wk of Navigator use during the first 4 weeks, which decreased slightly to 134 h/wk during weeks 9 to 13 (P = .006). Mean hemoglobin A1c improved from 7.1% at baseline to 6.8% at 13 weeks (P = .02), and the percentage of glucose values between 71 and 180 mg/dL increased from 52% to 60% (P = .01). Subjects and parents reported high satisfaction with the Navigator on the Continuous Glucose Monitor Satisfaction Scale. Two subjects had severe skin reactions related to sensor mount adhesive. This study indicates that incorporating real-time continuous glucose monitoring into the daily treatment of children with T1D is feasible. The results provide a compelling rationale for conducting a randomized trial of daily use of a continuous glucose monitor in children with T1D.
Medical subject headings
- Blood Glucose Self-Monitoring
- Diabetes Mellitus, Type 1
- Monitoring, Ambulatory