Computer-assisted total knee arthroplasty versus the conventional technique: how precise is navigation in clinical routine?
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 17899009.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Restoration of the mechanical leg axis and component positioning are crucial factors affecting long-term results in total knee arthroplasty (TKA). In a prospective study, 1,000 patients were operated on either using a CT-free navigation system or the conventional jig-based technique. Leg alignment and component orientation were determined on postoperative X-rays. The mechanical leg axis was significantly better in the computer-assisted group (95%, within +/-3 degrees varus/valgus) compared to the conventional group (74%, within +/-3 degrees varus/valgus) (P < 0.001). On average, the operating time was increased by 8 min in the computer-assisted group. No significant differences were seen between senior and younger surgeons regarding postoperative leg alignment and operating time. Computer-assisted TKA leads to a more accurate restoration of leg alignment and component orientation compared to the conventional jig-based technique. Potential benefits in long-term outcome and functional improvement require further investigation.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Joint Instability
- Surgery, Computer-Assisted
Anatomy
- knee