Radiation dose estimates in dual-source computed tomography coronary angiography.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 17909816.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to quantify radiation dose parameters of dual-source CT coronary angiography. Eighty patients underwent contrast-enhanced, retrospectively ECG-gated dual-source CT coronary angiography with heart rate-adapted ECG pulsing using two algorithms: In 40 patients, the tube current was reduced to 20% (A(min1)) of the normal tube current (A(max)) outside the pulsing window; in 40 patients tube current was reduced to 4% (A(min2)) of A(max). Mean CTDI(vol) in the A(min1) group was 45.1 +/- 3.6 mGy; the mean CTDI(vol) in the A(min2) group was 39.1 +/- 3.2 mGy, with CTDI(vol) in the A(min2) group being significantly reduced when compared to the A(min1) group (P < 0.001). A significant negative correlation was found between CTDI(vol) and heart rate in group A(min1) (r = -0.82, P < 0.001), whereas no correlation was found between CTDI(vol) and heart rate in group A(min2) (r = -0.066). Using the conversion coefficient for the chest, dual-source CT coronary angiography resulted in an estimated mean effective dose of 8.8 mSv in the A(min1) group and 7.8 mSv in the A(min2). Radiation exposure of dual-source CT coronary angiography using an ECG-pulsing protocol reducing the tube current to 20% significantly decreases with increasing heart rates, despite using wider pulsing windows at higher heart rates. When using a protocol with reduced tube current of 4%, the radiation dose is significantly lower, irrespective of the heart rate.
Medical subject headings
- Coronary Angiography
- Radiation Dosage
- Tomography, X-Ray Computed