In vitro analysis of exeter stem torsional stability.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 17920476.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The effect of cyclic loading on the torsional stiffness of a polished double-tapered femoral stem was investigated in vitro. Initial torsional stability was compared with torsional stability after cyclic loading. Stems were removed from the cement mantle and reinserted without the use of additional cement. Torsional stability was measured after reinsertion and after further cyclic loading. Subsidence of the stem was observed. No difference in torsional stiffness was observed during loading. No difference between the stiffness before extraction and after reinsertion was observed. Torsional stiffness of an Exeter stem does not decrease after axial subsidence under cyclic loading. Stability is retained after reinsertion into the original cement mantle. Debonding of the Exeter stem is not associated with rotational instability of the implant.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Torsion, Mechanical
Anatomy
- hip