Quality control in the conservative treatment of breast cancer: patient dosimetry using silicon detectors.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1792325.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Twenty patients with early breast cancer were treated with external irradiation, delivered with two tangential beams (6 MV X-rays) using a half-beam block (HBB) and 3-D compensating filters. All patients were immobilized with individualized cellulose acetate casts. Patient dosimetry was performed using p-type silicon detectors. Midline doses were calculated by combined entrance and exit dose measurements. The mean ratio of the measured and the prescribed doses was 96.6 +/- 3.8% at the reference point, 96.8 +/- 4.3% at off-axis points on the central plane and 96.8 +/- 7.6% at off-plane points.