Quality control in the conservative treatment of breast cancer: patient dosimetry using silicon detectors.

Ciocca, M; Landoni, L; Italia, C; Montanaro, P; Canesi, P; Valdagni, R · Radiother Oncol · 1991

case_series · Level IV

Where this comes from

Abstract

Twenty patients with early breast cancer were treated with external irradiation, delivered with two tangential beams (6 MV X-rays) using a half-beam block (HBB) and 3-D compensating filters. All patients were immobilized with individualized cellulose acetate casts. Patient dosimetry was performed using p-type silicon detectors. Midline doses were calculated by combined entrance and exit dose measurements. The mean ratio of the measured and the prescribed doses was 96.6 +/- 3.8% at the reference point, 96.8 +/- 4.3% at off-axis points on the central plane and 96.8 +/- 7.6% at off-plane points.