Test-retest reproducibility of elbow goniometric measurements in a rigid double-blinded protocol: intervals for distinguishing between measurement error and clinical change.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 17967547.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Increasingly, goniometry of elbow motion is used for qualification of research results. Expression of reliability is in parameters not suitable for comparison of results. We modified Bland and Altman's method, resulting in the smallest detectable differences (SDDs). Two raters measured elbow excursions in 42 individuals (144 ratings per test person) with an electronic digital inclinometer in a classical test-retest crossover study design. The SDDs were 0 +/- 4.2 degrees for active extension; 0 +/- 8.2 degrees for active flexion, both without upper arm fixation; 0 +/- 6.3 degrees for active extension; 0 +/- 5.7 degrees for active flexion; 0 +/- 7.4 degrees for passive flexion with upper arm fixation; 0 +/- 10.1 degrees for active flexion with upper arm retroflexion; and 0 +/- 8.5 degrees and 0 +/- 10.8 degrees for active and passive range of motion. Differences smaller than these SDDs found in clinical or research settings are attributable to measurement error and do not indicate improvement.
Medical subject headings
- Arthrometry, Articular
- Elbow Joint
- Range of Motion, Articular
Anatomy
- elbow