Thermal injury to the tongue from an operative laryngoscope.

McMullin, Brian T; Blumin, Joel H; Merati, Albert L · Otolaryngol Head Neck Surg · 2007

biomechanical · Level V

Where this comes from

Abstract

To understand risk and possible mechanisms of thermal injury to the tongue during suspension laryngoscopy. Tongue injury was noted following suspension laryngoscopy in a cluster of cases; the findings raised suspicion for thermal injury. To characterize the potential for thermal injury, a lighted laryngoscope was placed in contact with samples of tongue. Tissue heating was recorded over time at fixed distances from the light carrier within a laryngoscope. Comparisons were made between two light sources and carriers. Maximal tissue heating using a newer-model light source and new carrier was 16.7 degrees F immediately beneath the light carrier tip and 19.8 degrees F at 1 cm distal to the carrier tip. Other combinations of an older source and carriers failed to raise temperatures. Heating up to 10.7 degrees F occurred with the new source and an older carrier. Tissue heating may occur from contact with operative laryngoscopes. This is an important patient safety issue in laryngology.