Status of navigated total hip arthroplasty in dysplastic osteoarthritis.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 17983111.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This article describes the results of 41 total hip arthroplasties (THAs) carried out using a new image-free hip navigation system. For the study, registration landmarks (tear drop, posterior rim, and inner wall of the acetabulum) were added to the hip navigation system OrthoPilot (B. Braun Aesculap, Tuttlingen, Germany). The reamer center was indicated in reference to these landmarks, and final cup angle was indicated as the radiologic angle. Forty-one total THAs to treat dysplastic osteoarthritis of the hip were performed. The difference between reamer center and cup center was 4.1 +/- 3.0 mm. The difference in cup abduction angle between navigation and radiography was 6.4 degrees +/- 5.1 degrees and of anteversion was 5.9 degrees +/- 5.0 degrees. Our preliminary clinical results showed adequate accuracy.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Osteoarthritis, Hip
- Surgery, Computer-Assisted
Anatomy
- hip