Closed-loop feedback computer-controlled infusion of phenylephrine for maintaining blood pressure during spinal anaesthesia for caesarean section: a preliminary descriptive study.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 17991262.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We describe the novel use of a closed-loop feedback computer-controlled infusion of phenylephrine for maintaining blood pressure in 53 patients having spinal anaesthesia for elective caesarean section. A simple on-off algorithm was used that activated an intravenous phenylephrine infusion at 100 microg.min(-1) when systolic blood pressure was less than or equal to baseline and stopped the infusion when systolic blood pressure exceeded baseline. Up to uterine incision, 94.6% of all systolic blood pressure measurements were within the range (baseline +/- 20%). Seven patients (13.2%) had one or more episodes of hypotension (systolic blood pressure < 80% of baseline) and 23 patients (37.7%) had one or more episodes of hypertension (systolic blood pressure > 120% of baseline). No patient had nausea or vomiting and in no case was umbilical arterial blood pH < 7.2. Calculated system performance parameters were comparable with those of previously published closed-loop systems and provide a reference for the potential development and comparison of more advanced algorithms.
Medical subject headings
- Anesthesia, Obstetrical
- Anesthesia, Spinal
- Cesarean Section
- Hypotension
- Phenylephrine
- Vasoconstrictor Agents