Intermediate results of a cemented femoral stem with a PMMA premantle.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 18019989.
- Also identified by DOI 10.3928/01477447-20071101-11.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This retrospective study evaluated the intermediate clinical results ot a series of 143 patients after total hip arthroplasty using the Definition PM femoral component. One hundred twenty-two patients had a mean follow-up of 5 years (range, 23-77 months), died (mean 30 months; range, 6-56 months), or were revised (mean 20 months; range, 9-40 months). The mechanical failure rate of the femoral stem was 0.8% for those patients with an average 5-year follow-up. Radiographic examination revealed that 1 (0.8%) femoral component was possihly loose, while none had evidence of dehonding. Follow-up radiographs revealed 2.50% (anteroposterior) and 5% (lateral) of the femoral components to have radiolucencies at the cement-bone interface. At a mean of 5 years. 2 of the 122 femoral romnonents imnlanted renuired revision.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Bone Cements
- Hip Prosthesis
Anatomy
- femur
- hip