Bilateral ce-MR angiography of the hands at 3.0 T and 1.5 T: intraindividual comparison of quantitative and qualitative image parameters in healthy volunteers.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 18040693.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to determine the benefit of bilateral contrast-enhanced MR angiography (ce-MRA) of the hands at 3.0 Tesla (T) compared with an established 1.5-T technique in healthy volunteers. Intraindividual bilateral ce-MRA of the hands was performed at 1.5 T and 3.0 T in 14 healthy volunteers using a timed ultra-fast GRE sequence featuring parallel acquisition. The evaluation comprised measurement of the vessel signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), rating of the image quality and the assessment of artefacts and venous contamination. At 3.0 T, SNR improved up to 95% and CNR up to 129%. The image quality of the larger inflow arteries, the palm arches and common digital arteries was good or sufficient at either magnetic field strengths. However, 3.0-T MRA was clearly superior in the depiction of the digital arteries. Ce-MRA of the hand clearly profits from the use of 3.0 T. Compared with 1.5 T, a substantial increase of CNR is found resulting in a significantly better delineation of the small digital arteries. Saturation affects more the SNR of the perivascular tissue than the contrast-enhanced blood, and thus leads to a marked increase of CNR at 3.0.
Medical subject headings
- Hand
- Magnetic Resonance Angiography