The Humphrey ADE system: evaluation in paediatric use.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 1817632.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Humphrey ADE single lever parallel form breathing system was evaluated in 17 children whose lungs were ventilated mechanically, mean age 28 (range 9-58) months, weight 13.3 (8.3-18.0) kg and in 10 spontaneously breathing children, age 50.7 (18-99) months and weight 17.5 (10.9-24.3) kg. During controlled ventilation in the E mode, the ADE behaved in a manner similar to the Jackson-Rees modification of Ayre's T-piece with the fresh gas flows (FGF) set either according to weight or at a standard 3 litre min-1. After 5 min hand ventilation in the A mode with FGF 3 litre min-1 the mean end-tidal partial pressure of carbon dioxide was 5.3 (SD 0.8) kPa (range 4.1-7.7 kPa). Whilst breathing spontaneously in the A mode, no patient experienced rebreathing at FGF 3 litre min-1; rebreathing started at 124 (31.6) ml kg-1 min-1. The ADE circuit performed satisfactorily in all three modes. During controlled ventilation in the E mode, FGF should be adjusted according to the standard formulae recommended for the T-piece circuit. During hand ventilation in the A mode, an FGF greater than 3 litre min-1 should be used.
Medical subject headings
- Anesthesia, Closed-Circuit
- Pediatrics