Oxidation and wear of 100-Mrad cross-linked polyethylene shelf-aged for 30 years.

Oonishi, Hironobu; Kim, Sok Chol; Oonishi, Hiroyuki; Kyomoto, Masayuki; Masuda, Shingo · Clin Orthop Relat Res · 2008

biomechanical · Level V

Where this comes from

Abstract

Some previous studies suggest that aging influences wear and oxidatively degraded nonsterilized ultra-high-molecular-weight polyethylene (UHMWPE) exhibits decreased wear resistance. We therefore asked whether shelf-aging storage conditions influenced degradation and wear resistance of gamma-irradiated UHMWPE. We examined oxidation and wear of 100-Mrad gamma-irradiated UHMWPE (100-Mrad polyethylene) cups shelf-aged for 30 years without (n=2) or with (n=2) packages. The oxidation index of the unpackaged 100-Mrad polyethylene surface (4) was higher than that of the packaged one (2.7). The packaged 100-Mrad polyethylene cup exhibited a high wear resistance with a steady wear rate of 0.5 mg/10(6) cycles. In contrast, the unpackaged 100-Mrad polyethylene exhibited an extremely high initial wear rate of 187.9 mg/10(6) cycles over the first 0.25 x 10(6) cycles with a subsequently reduced wear rate of 5 mg/10(6) cycles after 5 x 10(6) cycles. Packaging over long periods inhibits surface oxidation and maintains the wear resistance of gamma-irradiated UHMWPE cups.

Medical subject headings

Anatomy