Suppression of slice selection axis motion artifacts in MRI.

Hedley, M; Yan, H · IEEE Trans Med Imaging · 1992

basic_science · Level V

Where this comes from

Abstract

Patient motion during data acquisition in magnetic resonance imaging causes artifacts in the reconstructed image, which for two-dimensional Fourier transform imaging techniques appear as blurring and ghost repetitions of the moving structures. T. Mitsa et al. (1990) proposed a technique for suppressing artifacts from periodic motion along the slice selection axis. A different approach to the same problem is presented which is not restricted to periodic motion. The algorithm is verified using a simulated phantom and motion. It is also shown to perform well in the presence of noise and motion within the imaging plane.