Comments on ;A cone-beam filtered backprojection reconstruction algorithm for cardiac single photon emission computed tomography' by G. T. Gullberg and G. L. Zeng.
editorial · Level V
Where this comes from
- Record sourced from PubMed, PMID 18218429.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In the above-titled work by G. T. Gullberg and G. L. Zeng (ibid., vol.11, no.1, p.91-101, 1992), a fan-beam reconstruction formula of a noncircular scanning locus was derived and extended for half-scan cone-beam reconstruction. However, the reconstruction formula is not exact mathematically unless a necessary condition is satisfied. In this correspondence, the commenters derive this necessary condition and provide a geometrical explanation of the condition.