Evaluation of a new vascular suture system for aortic laparoscopic surgery: an experimental study on pigs and cadavers.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 18261940.
- Also identified by DOI 10.1016/j.ejvs.2007.12.010.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to assess the feasibility and efficacy of a new laparoscopic vascular suturing device. Animal study: six pigs underwent surgery using a retroperitoneal laparoscopic approach. Aorto-prosthetic side-to-end and end-to-end anastomoses were performed laparoscopically on each pig using SuDyn. Clamping and anastomosis times, as well as the properties of the anastomoses, were recorded. Study on cadavers: four aorto-prosthetic end-to-end anastomoses were performed using the direct transperitoneal laparoscopic approach to assess the feasibility of the SuDyn device on atherosclerotic aortas. Animal study: No pigs died and 12 patent and impermeable anastomoses were obtained. Mean anastomosis time was 38(+/-8)min for end-to-side anastomoses and 37(+/-5)min for end-to-end anastomoses. Study on cadavers: Totally laparoscopic anastomoses were performed in 4 human cadavers with a mean anastomosis time of 37(+/-3)min. SuDyn makes laparoscopic aorto-prosthetic anastomoses easier to perform, produces good results and does not require a learning curve.
Medical subject headings
- Aorta
- Blood Vessel Prosthesis Implantation
- Laparoscopes
- Laparoscopy
- Suture Techniques