Hierarchy of stability factors in reverse shoulder arthroplasty.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 18264855.
- Also identified by DOI 10.1007/s11999-007-0096-0 and PMC identifier 2505225.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Reverse shoulder arthroplasty is being used more frequently to treat irreparable rotator cuff tears in the presence of glenohumeral arthritis and instability. To date, however, design features and functions of reverse shoulder arthroplasty, which may be associated with subluxation and dislocation of these implants, have been poorly understood. We asked: (1) what is the hierarchy of importance of joint compressive force, prosthetic socket depth, and glenosphere size in relation to stability, and (2) is this hierarchy defined by underlying and theoretically predictable joint contact characteristics? We examined the intrinsic stability in terms of the force required to dislocate the humerosocket from the glenosphere of eight commercially available reverse shoulder arthroplasty devices. The hierarchy of factors was led by compressive force followed by socket depth; glenosphere size played a much lesser role in stability of the reverse shoulder arthroplasty device. Similar results were predicted by a mathematical model, suggesting the stability was determined primarily by compressive forces generated by muscles.
Medical subject headings
- Arthroplasty, Replacement
- Joint Instability
- Rotator Cuff
- Shoulder Dislocation
- Shoulder Joint
Anatomy
- shoulder