A new geometric factor for in situ resistivity measurement using four slender cylindrical electrodes.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 18269995.
- Also identified by DOI 10.1109/TBME.2007.903704.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The four-electrode method is commonly used for in situ measurement of the electrical resistivity of biological tissues. In this paper, a new geometric factor between the resistivity and measured resistance using the four-electrode interface is derived in the prolate spheroidal coordinates and experimentally validated. Evaluation of the experimental results shows that the resistivities determined using both the derived geometric factor and a commercial conductivity meter are in close agreement even when the length of the immersed electrodes becomes long with respect to the inter-electrode spacing. The evaluation also shows the effect of the relative size of the sample volume when the limitation to semi-infinite volume begins to result in poor accuracy.
Medical subject headings
- Algorithms
- Electrodes
- Models, Biological
- Plethysmography, Impedance