A pitfall of transfix fixation during anterior cruciate ligament reconstruction.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 18270684.
- Also identified by DOI 10.1007/s00167-008-0489-y.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
While attempting femoral fixation during anterior cruciate ligament (ACL) reconstruction using Bio-TransFix (Arthrex, Naples, FL), we experienced failure of complete insertion of a hamstring graft into the femoral tunnel. A passing wire was twisted in the femoral tunnel during insertion of the ACL graft. The cause and methods of prevention of this complication are discussed.
Medical subject headings
- Anterior Cruciate Ligament
- Intraoperative Complications
- Orthopedic Fixation Devices
- Tendons
Anatomy
- femur