Dislocation of rotating-hinge knee prostheses with antidislocation mechanism.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 18280428.
- Also identified by DOI 10.1016/j.arth.2006.11.020.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
At our institution, we have used Endo-Model prostheses (Waldermar Link GmbH & Co, Hamburg, Germany) for difficult knee reconstructions. These rotating-hinge prostheses have an antidislocation feature that prevents dislocation due to distraction disengagement. In a series of 192 consecutive prostheses, we have identified 2 cases of femorotibial dislocations. An increased flexion laxity allows excessive distraction and hence implant dislocation after antidislocation feature loosening. To our knowledge, there are no other reports in the literature regarding dislocation of a rotating-hinge knee prosthesis with antidislocation design.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Dislocation
- Knee Prosthesis
- Prosthesis Failure
Anatomy
- knee