An investigation into the length of standard tracheostomy tubes in critical care patients.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 18289238.
- Also identified by DOI 10.1111/j.1365-2044.2007.05327.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A number of problems have been reported with the use of standard length tracheostomy tubes in adult critical care patients. We measured the depth from the skin surface to the tracheal wall and the angle of the tracheal stoma during percutaneous tracheostomy. In vitro measurements were also performed on standard tracheostomy tubes. Comparison of in vivo and in vitro dimensions demonstrated that standard length tracheostomy tubes are too short for the average critical care patient. We recommend that both the stomal and intratracheal lengths should be made longer by approximately 1 cm and tubes should be redesigned to an angle of 110-120 degrees to allow optimal tracheal placement.
Medical subject headings
- Critical Care
- Tracheostomy