Cardiac output measurement in children: comparison of the Ultrasound Cardiac Output Monitor with thermodilution cardiac output measurement.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 18297271.
- Also identified by DOI 10.1007/s00134-008-1030-y.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare the assessment of cardiac output (CO) in children using the noninvasive Ultrasound Cardiac Output Monitor (USCOM) with the invasive pulmonary artery catheter (PAC) thermodilution cardiac output measurement. Prospective observational study in a tertiary center for pediatric cardiology of a university children's hospital. Twenty-four pediatric patients with congenital heart disease without shunt undergoing cardiac catheterization under general anesthesia. CO was measured by USCOM using a suprasternal CO Doppler probe in children undergoing cardiac catheterization. USCOM data were compared to CO simultaneously measured by PAC thermodilution technique. Measurements were repeated three times within 5 min in each patient. A mean percentage error not exceeding 30% was defined as indicating clinical useful reliability of the USCOM. CO values measured by PAC ranged from 1.3 to 5.3 l/min (median 3.6 l/min). Bias and precision were -0.13 and 1.34 l/min, respectively. The mean percentage error of CO measurement by the USCOM compared to PAC thermodilution technique was 36.4% for USCOM. Our preliminary data demonstrate that cardiac output measurement in children using the USCOM does not reliably represent absolute CO values as compared to PAC thermodilution. Further studies must evaluate the impact of incorporating effective aortic valve diameters on CO measurement using the USCOM.
Medical subject headings
- Cardiac Output
- Heart Defects, Congenital
- Monitoring, Physiologic
- Thermodilution
- Ultrasonography