Vitreous dynamics: vitreous flow analysis in 20-, 23-, and 25-gauge cutters.

Magalhaes, Octaviano; Chong, Lawrence; DeBoer, Charles; Bhadri, Prashant; Kerns, Ralph; Barnes, Aaron; Fang, Sophia; Humayun, Mark · Retina · 2008

basic_science · Level V

Where this comes from

Abstract

To evaluate porcine vitreous flow and balanced saline solution (BSS) flow rates in different vitrectomy systems. Porcine vitreous was obtained within 24 hours of slaughter. A high-speed (2 samples/s) balance, precise to 0.01 g, was used. Variable cut rates and vacuum pressures were analyzed in vitreous and BSS. The vitreous was labeled with glass microspheres and triamcinolone acetonide. A high-speed (400 frames/s) camera was used to record cutting for each condition. For all cutters, there was no vitreous flow at zero cut rates (off). In 25-gauge cutters, at 500 mmHg of vacuum, the electric cutter produced higher average flow rates at high cut rates (600 cpm, 0.004 mL/s, and 1500 cpm, 0.013 mL/s) than pneumatic, which demonstrated a decreased flow at speeds higher than 1000 cpm (1000 cpm, 0.015 mL/s, and 1500 cpm, 0.006 mL/s). The percentage of vitreous flow rate/BSS flow rate in different aspiration and cut rates showed an ascending curve. This demonstrates evidence of flow obstruction in 25- (all cut rates), 23- (all cut rates), and 20-gauge (all cut rates). Flow obstruction and surge movements were seen in the high-speed videos. The vitrectomy systems each illustrate different performances of vitreous removal. The physical characteristics of vitreous resulted in nonuniform flow in all vitreous cutters.