A computerized analysis-by-synthesis algorithm improves precision of linear wear measurements in total hip replacements.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 18327802.
- Also identified by DOI 10.1002/jor.20643.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Precision is important if small polyethylene wear rates are to be detected early and in small sample sizes. Using an automatic, computerized algorithm relying on a synthetic X-ray generated from a computer-assisted design (CAD) model of the implant may significantly improve precision of linear wear measurements. We compared this method to provide improved in vivo precision compared to the widely used Hip Analysis Suite software (version 8.0.1.7). Two anterior-posterior pelvic X-rays were taken on the same day of 18 total hip implants, and wear was measured by three observers using the two methods. Expected real wear was zero and defined as a reference value. Two sources of variability were estimated: one due to the instrument, and one due to different operators. A Wilcoxon Signed Rank Test was used for statistical comparisons. The overall precision (instrument variability and multioperator variability) given a 95% precision limit was significantly higher (p < 0.001) using the proposed method (+/-0.15 mm) when compared to the Hip Analysis Suite (+/-0.84 mm). Due to the availability of the full CAD data of the implant and to further automation of the measuring process, the newly proposed method provided improved precision in measuring polyethylene wear.
Medical subject headings
- Arthrography
- Arthroplasty, Replacement, Hip
- Equipment Failure Analysis
- Hip Prosthesis
- Prosthesis Failure
Anatomy
- hip