Early experience with the TriMed fragment-specific fracture fixation system in intraarticular distal radius fractures.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 18332021.
- Also identified by DOI 10.1177/1753193407087887.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper reports the results of fixation of intraarticular fractures of the distal radius using the TriMed fragment-specific fixation system by surgeons early in their experience with the system. A consecutive series of 22 AO types C2 and C3 fractures in 21 patients were internally fixed with the system. Restoration of articular congruity to less than 2 mm was possible in 20 fractures. At a minimum of 6 months follow-up, mean volar and radial inclination were 8 degrees and 25 degrees , respectively, with no loss of reduction. The patients had a mean of 50 degrees flexion, 63 degrees extension and a pronation-supination arc of 149 degrees . The mean subjective Patient Rated Wrist Score was 20. Our complication rate was comparable to other published series despite long operating times. Therefore, we commend the system as a powerful tool to treat these difficult fractures but acknowledge the significant learning curve.
Medical subject headings
- Fracture Fixation, Internal
- Fractures, Comminuted
- Radius Fractures
Anatomy
- radius