A quantitative method based on total relative change for dynamic electrical impedance tomography.

You, Fusheng; Shi, Xuetao; Dong, Xiuzhen; Fu, Feng; Liu, Ruigang; Shuai, Wanjun; Li, Zheng · IEEE Trans Biomed Eng · 2008

basic_science · Level V

Where this comes from

Abstract

We proposed a new method based on total relative change (TRC) from measured boundary voltages to quantify the volume changes of fluid during electrical impedance tomography (EIT) monitoring. The results showed that TRC linearly correlated with the volume of infused saline solution into a phantom, and the slope of TRC changes was approximately linear with the infusion speed. A inserted copper tube at different positions did not affect TRC significantly. The linear relationship between TRC and volume change indicates that TRC could be a good quantitative index for dynamic EIT.