Laryngoscopy force in the sniffing position compared to the extension-extension position.
other
Where this comes from
- Record sourced from PubMed, PMID 18336487.
- Also identified by DOI 10.1111/j.1365-2044.2008.05492.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Laryngoscopy is sometimes easier with the patient's head and neck in the extension-extension position (head extension with the neck extended by the head section of the table bent down at 30 degrees) rather than the classical 'sniffing the morning air' position. We therefore tested the hypothesis that the axial force required for laryngoscopy is less in the extension-extension than the sniffing position. We measured the force axial to the handle of a Macintosh 3 laryngoscope in 20 subjects under general anaesthesia who had been given neuromuscular blocking drugs. Measurement of force was made in the sniffing position and the extension-extension position. The mean (SD) axial force required in the extension-extension position was lower than in the sniffing position (19.6 (7.8) N versus 23.6 (8.6) N, p = 0.04). In the setting of routine tracheal intubation, less force is required when the patient is in the extension-extension position than in the sniffing position.
Medical subject headings
- Head
- Laryngoscopy
- Neck