Biomechanical characteristics of nonbridging external fixators for distal radius fractures.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 18343285.
- Also identified by DOI 10.1016/j.jhsa.2007.09.019.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Nonbridging external fixation is becoming popular for distal radius fractures, although its biomechanical characteristics have not been documented. This study evaluated the biomechanical characteristics of nonbridging external fixators for distal radius fractures. We tested 3 currently available nonbridging fixators (F-Wrist fixator, Hoffman II Compact, and Pennig Dynamic Wrist Fixator) and determined their relative stiffness under axial compression, torsion, and bending moments (dorsal, volar, radial, and ulnar aspects) using a uniform unstable distal radius fracture model. The contact pressure and its total load on the fracture plane were also measured to evaluate the mechanical stimuli at the stable fracture site using a pressure-sensitive conductive rubber sensor. Differences were observed in the stiffness: the Pennig fixator was the stiffest, whereas the F-Wrist fixator was the least rigid. The total load transmitted from the wrist joint to the fracture plane depended on the fixator stiffness in axial compression. By determining the biomechanical characteristics of nonbridging external fixators, these data may help the clinician when deciding on a particular device for nonbridging external fixation.
Medical subject headings
- External Fixators
- Radius Fractures
Anatomy
- radius