Analytic bias among certified methods for the measurement of hemoglobin A1c: a cause for concern?
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 18343780.
- Also identified by DOI 10.1309/U3GPPTCBP1VLL8AW.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We studied the magnitude, significance, and origin of an analytic bias that emerged between our point-of-care (POC) and our central laboratory (CL) methods for the measurement of hemoglobin A1c (HbA1c) and evaluated the analytic accuracy of 7 commonly used HbA1c methods relative to the National Glycohemoglobin Standardization Program (NGSP) reference method. The POC and CL methods were compared by split-sample analysis of clinical specimens and time series analyses of the HbA1c results reported for a 33-month period. The relative accuracies of 7 HbA1c methods were evaluated using College of American Pathologists proficiency survey results. Long-term drifts in the CL- and POC-analyzed test results caused the median intermethod bias [(POC result)-(CL result)] to increase from -0.4% to -0.9% HbA1c. Systematic biases, drifts in analytic performance over time, and intermethod variability were frequently observed among the 7 NGSP-certified HbA1c methods. Intermethod variability is a potential source of inaccuracy whenever HbA1c results are interpreted relative to universal, fixed, clinical decision thresholds.
Medical subject headings
- Blood Chemical Analysis
- Glycated Hemoglobin
- Laboratories
- Quality Control