Bridge to lung transplantation through a pulmonary artery to left atrial oxygenator circuit.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 18355495.
- Also identified by DOI 10.1016/j.athoracsur.2007.12.032.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
There is no mechanical device available to support patients with end-stage lung failure for weeks and months until appropriate donor organs for lung transplantation are available. In a 38-year-old female patient with primary pulmonary hypertension a paracorporeal artificial lung (PAL) system was placed parallel to the pulmonary circulation with connections to the pulmonary artery and to the left atrium. The key component of the PAL was a low-resistance membrane oxygenator. After institution, the PAL had a blood flow of 3.5 L/min and created a PaO(2)/fraction of inspired oxygen ratio of 270, while the oxygenator was provided with oxygen 3 L/min. The pulmonary artery pressure declined by almost 50%. The PAL worked well over 62 days until appropriate donor lungs were available. With resuming more physical activity, an increased flow through the native lung augmented the fraction of unsaturated blood arriving at the left atrium, which mandated increasing oxygen flow to the PAL. The data obtained with this case encourage further research into PAL systems, which may hopefully serve as a bridge to lung transplant device in appropriate patients in the future.
Medical subject headings
- Artificial Organs
- Atrial Function, Left
- Hypertension, Pulmonary
- Lung
- Oxygenators
- Pulmonary Circulation