Use of the HemoCue near patient testing device to measure the concentration of haemoglobin in suction fluid at elective caesarean section.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 18412653.
- Also identified by DOI 10.1111/j.1365-2044.2007.05400.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We have investigated the suitability of the HemoCue photometer to measure the concentration of haemoglobin in suction fluid obtained at elective caesarean section in 30 women. Laboratory analysis was used as a gold standard against which values generated by the HemoCue were compared. We used the method of Bland and Altman to analyse the data. The bias and the limits of agreement were -0.013 and -0.39 to 0.36 mg x dl(-1) respectively, indicating a good level of agreement. Mean (SD) total blood loss calculated using these data, combined with the weight of the swabs, was consistently greater than clinical estimation: 768 (496) ml versus 506 (249) ml respectively (p < 0.001). We have found that the HemoCue near patient testing device may be used to estimate blood loss accurately in the suction fluid obtained at elective Caesarean section.
Medical subject headings
- Blood Loss, Surgical
- Cesarean Section
- Hemoglobinometry
- Point-of-Care Systems