The Amplatzer vascular plug for large vessel occlusion in the endovascular management of aneurysms.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 18418603.
- Also identified by DOI 10.1007/s00330-008-0967-0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Amplatzer vascular plug (AVP) is derived from the Amplatzer device used in correction of cardiac septal defects. We present a large series of the use of the AVP in the endovascular management of aneurysms. Three patients with a combination of aortic and peripheral aneurysms underwent embolisation with the AVP. Plugs with 10-16 mm diameter were used and delivered using introducer/guiding sheaths. A total of 16 internal iliac arteries, 2 common iliac arteries (CIA), 4 subclavian (SCA) arteries, 1 superior mesenteric (SMA) and 1 popliteal artery were embolised. Successful occlusion with the AVP was achieved in 21 out of 24 vessels (87.5%), of which 18 (75%) occluded immediately and 3 (12.5%) were delayed occlusions. The three patients who are considered to have failed to occlude with the AVP required the use of additional embolic agents. At the end of the 17-month follow-up all 24 target vessels had occluded. Two patients developed persistent buttock claudication, and one had a left hemiparesis. The AVP is useful as an adjunctive device in the management of aneurysms. It has a particular role in embolisation of large-diameter vessels with a short implantation zone. The device is safe and easy to use.
Medical subject headings
- Aneurysm
- Blood Vessel Prosthesis
- Embolization, Therapeutic