Cutaneous burns from a fiberoptic cable tip during arthroscopy of the knee.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 18514527.
- Also identified by DOI 10.1016/j.knee.2008.02.007.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Thermal skin injury is a rare complication of arthroscopy. The potential of fiberoptic light cable tips to cause cutaneous burns has been previously investigated, but no cases have been reported. We present a case report of such an injury. In order to improve patient safety, we further recommend always checking that the light source is set on standby prior to disconnecting the light cable from the arthroscope and that the light cable tip is never left on the patient when not in use.
Medical subject headings
- Anterior Cruciate Ligament Injuries
- Arthroscopy
- Burns
- Fiber Optic Technology
- Knee Injuries
Anatomy
- knee