Tibial post failures in a condylar posterior cruciate substituting total knee arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 18534385.
- Also identified by DOI 10.1016/j.arth.2007.08.002.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In posterior-stabilized total knee arthroplasties, a femoral cam and polyethylene tibial post are commonly used to restore posterior stability after sacrifice of the posterior cruciate ligament. This article reports a high incidence of early tibial post failures in one design of prosthesis and examines the variables that may have contributed to such. Five hundred sixty-four consecutive posterior-stabilized total knees were implanted in 512 patients, using a total knee prosthesis with a polyethylene tibial post and femoral cam. Clinical and radiographic outcomes were measured at a mean follow-up of 40 months after surgery (range, 24-83 months). At follow-up, 70 knees in 62 patients (12%) had undergone revision surgery because of symptoms related to catastrophic failure of the tibial post.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
- Posterior Cruciate Ligament
- Prosthesis Design
- Tibia