Effects of the femoral offset and the head size on the safe range of motion in total hip arthroplasty.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 18534445.
- Also identified by DOI 10.1016/j.arth.2008.02.008.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to quantify the effects of femoral offset and head size on range of motion (ROM) after total hip arthroplasty. Modular prostheses were implanted into 11 cadaveric hips using a posterolateral approach and tested for ROM with 3 different offsets and 5 different femoral head sizes. Increasing the femoral offset to 4 and 8 mm resulted in 21.1 degrees and 26.7 degrees of improved flexion, and 13.7 degrees and 21.2 degrees of improved internal rotation, respectively. The ROM improved in a head size-dependent manner primarily because of increasing the jumping distance of the femoral head rather than delaying any impingement. In contrast, the effectiveness of femoral offset was driven by delayed osseous impingement.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Joint
- Hip Prosthesis
- Range of Motion, Articular
Anatomy
- femur
- hip