Measurement of the acetabular cup anteversion on simulated radiographs.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 18534457.
- Also identified by DOI 10.1016/j.arth.2007.10.029.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Widmer (J Arthroplasty 2004;19:387) reported a protractor for measuring the anteversion of acetabular cups on radiographs but with limited precision. We intended to improve its precision by trigonometric mathematics. We measured the anteversion of the acetabular cups on 336 simulated radiographs using aforementioned 2 methods. The anteversion measured by Widmer's protractor ranged from 7 degrees to 41 degrees (mean +/- SD = 28.0 degrees +/- 9.8 degrees), and our methods, 5 degrees to 51 degrees (27.7 degrees +/- 13.2 degrees). The mean +/- SD of error by Widmer's protractor was 5.2 +/- 2.5 degrees, and our protractor, 0.8 degrees +/- 0.8 degrees (Student t test, P b .0001). The interobserver study showed the difference between measurements less than 2 degrees for each method. Therefore, the smaller error of our method than that of Widmer implicated a potentially precise measurement of the anteversion (level of evidence: diagnostic study, level II).