Surface damage after multiple dislocations of a 38-mm-diameter, metal-on-metal hip prosthesis.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 18534503.
- Also identified by DOI 10.1016/j.arth.2007.09.007.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We present, for the first time, a detailed damage assessment of a large-diameter metal-on-metal (L-MOM) hip prosthesis to show the extent of surface damage that can occur in a patient after multiple dislocations. The patient was a man (51 years old) who dislocated 8 times and was finally revised at 27 months. Radiographically, the cup was malpositioned with 65 degrees lateral opening and 15 degrees retroversion. The retrieved cup was a 1-piece, 38-mm Co-Cr-Mo (M2a; Biomet, Warsaw, Ind) with a titanium-alloy backing. The retrieved components demonstrated all known modes of wear, including a polished wear scar, multidirectional scratching, "stripe" wear, surface contamination of titanium-alloy, front face wear, and backside wear. The clinical significance is that cup positioning remains critical regardless of whether a large diameter head is used or not.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Dislocation
- Hip Prosthesis
- Prosthesis Failure
Anatomy
- hip