Stability differentials for proximal vs distal fusion of total hip arthroplasty femoral impaction grafts.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 18534514.
- Also identified by DOI 10.1016/j.arth.2007.08.006 and PMC identifier 2604125.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In impaction grafting for revision joint arthroplasty, the morselized cancellous bone ideally remodels into a new contiguous lattice. However, the use of biologically active factors may sometimes be indicated to enhance fusion. The purpose of this study was to determine the stability of femoral impaction-graft constructs for which either only the proximal or distal half of the morselized cancellous bone volume was modeled as fused. Fusing the proximal half of the impaction-graft volume resulted in a higher femoral stem stability than did fusing the distal half. This proximal graft fusion also resulted in a stem stability that was similar to that of fusing the entire graft. These results emphasize the importance of proximal fixation of an impaction-grafted femoral stem.