Proton beam therapy interference with implanted cardiac pacemakers.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 18538490.
- Also identified by DOI 10.1016/j.ijrobp.2008.01.062.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To investigate the effect of proton beam therapy (PBT) on implanted cardiac pacemaker function. After a phantom study confirmed the safety of PBT in patients with cardiac pacemakers, we treated 8 patients with implanted pacemakers using PBT to a total tumor dose of 33-77 gray equivalents (GyE) in dose fractions of 2.2-6.6 GyE. The combined total number of PBT sessions was 127. Although all pulse generators remained outside the treatment field, 4 patients had pacing leads in the radiation field. All patients were monitored by means of electrocardiogram during treatment, and pacemakers were routinely examined before and after PBT. The phantom study showed no effect of neutron scatter on pacemaker generators. In the study, changes in heart rate occurred three times (2.4%) in 2 patients. However, these patients remained completely asymptomatic throughout the PBT course. PBT can result in pacemaker malfunctions that manifest as changes in pulse rate and pulse patterns. Therefore, patients with cardiac pacemakers should be monitored by means of electrocardiogram during PBT.
Medical subject headings
- Cardiac Pacing, Artificial
- Heart Block
- Neoplasms
- Pacemaker, Artificial
- Proton Therapy
- Radiotherapy