Delayed catastrophic failure of a ceramic head in hybrid total hip arthroplasty.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 18538535.
- Also identified by DOI 10.1016/j.arth.2008.01.310.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Catastrophic failure of ceramic implants is a rare but known complication of total hip arthroplasty where such implants have been used. We report an unusual case of a 67-year-old woman who underwent bilateral hybrid total hip arthroplasty using ceramic-on-polyethylene bearings. Seven years later, she presented with left-sided hip pain in the absence of any trauma. She was found to have a completely shattered ceramic head, with subsequent erosion of the metal taper of the femoral stem through the polyethylene insert, its metal shell, and into the pelvis. This complicated subsequent revision surgery, as removal of the acetabular component resulted in a large pelvic defect that required an acetabular reinforcement ring and bone grafting. She recovered well after the revision procedure.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Ceramics
- Hip Prosthesis
- Prosthesis Failure
Anatomy
- hip