Blunt trauma-induced pacemaker failure.

Brown, K R; Carter, W; Lombardi, G E · Ann Emerg Med · 1991

case_report · Level V

Where this comes from

Abstract

A 54-year-old man with an artificial pacemaker sustained blunt trauma to his chest when he was struck with a baseball bat. Within 15 minutes after the injury, the patient experienced cardiovascular collapse. His pacemaker failed, and he required insertion of a temporary transvenous pacemaker. At surgery, the defect was traced to failure of the pulse generator, a rare cause of pacemaker failure. Emergency department evaluation should include prompt and continuous ECG monitoring, an overpenetrated chest radiograph, and telemetry evaluation after discharge.