Acute liner disassociation of a Pinnacle acetabular component.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 18555645.
- Also identified by DOI 10.1016/j.arth.2008.03.010.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A primary total hip arthroplasty using a cementless Pinnacle cup and neutral Marathon polyethylene liner mated via a 32-mm cobalt chrome head to a Summit pore coated stem (Depuy, Warsaw, IN). At 23 months postoperative an acute disassociation occurred. This was resolved with a liner exchange to a 36-mm metal liner with a 36-mm metal head. The retrieved polyethylene liner was noted to have 3 of the 6 peripheral locking tabs sheared off.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Osteoarthritis, Hip
- Prosthesis Design
- Prosthesis Failure