False-negative capnographic reading caused by a malfunctioning bag-valve-mask device resulting in a pneumomediastinum.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 18556108.
- Also identified by DOI 10.1016/j.resuscitation.2008.04.003.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Bag-valve-mask ventilation is generally a safe, and often a life-saving procedure for children who are unable to oxygenate or ventilate by themselves. A case is presented which describes a malfunctioning expiratory valve in a disposable manual bag-valve-mask device resulting in a pneumomediastinum in a child. This is a rare cause of a false-negative colorimetric end-tidal carbon dioxide (ETCO2) reading after endotracheal intubation. However, it is important to recognize so that further adverse outcomes can be prevented.
Medical subject headings
- Capnography
- False Negative Reactions
- Masks
- Mediastinal Emphysema