Comparison of cardiac output measurement using the CardioQP oesophageal Doppler with cardiac output measurement using thermodilution technique in children during heart catheterisation.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 18557743.
- Also identified by DOI 10.1111/j.1365-2044.2008.05495.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The minimally invasive CardioQP oesophageal Doppler probe estimates cardiac output by measuring blood flow velocity in the descending aorta. Individual variables to enter are patient's age, weight and height. We measured cardiac output simultaneously with CardioQP and pulmonary artery catheter thermodilution techniques during heart catheterisation in 40 paediatric patients with congenital heart defects. Median [range] age was 8.2 years [0.5-16.7 years], cardiac output values measured by thermodilution and CardioQP were 3.6 l.min(-1) [1.2-7.1 l.min(-1)] and 3.0 l.min(-1) [0.7-6.7 l.min(-1)], respectively. These values showed only moderate correlation (r = 0.809; p < 0.0001). Bias and precision were 0.66 l.min(-1) and 1.79 l.min(-1) (95% limits of agreement: -1.13 to +2.45 l.min(-1)). Based on our preliminary experience, cardiac output values measured by CardioQP in children do not reliably represent cardiac output values compared with the thermodilution technique. We suggest measurement of individual aortic diameter to improve performance of the CardioQP.
Medical subject headings
- Cardiac Catheterization
- Cardiac Output
- Heart Defects, Congenital