A beamformer for the acquisition of evoked potentials.

McKinley, C A; Parker, P A · IEEE Trans Biomed Eng · 1991

basic_science · Level V

Where this comes from

Abstract

Evoked potentials (EP) contain information about various physiological parameters and the estimation and detection of these signals can aid in the diagnosis of many pathological conditions. However, the signal-to-noise ratio (SNR) for EP measurement is often very low, and thus signal processing techniques must be employed to enhance the SNR. A delay and sum beamformer acquisition system has the potential for significant SNR improvement in EP measurements. In this communication it is shown that an electrode array acquisition system implements a uniform coherent delay and sum beamformer. The performance of the beamformer is characterized in terms of the number of electrodes, and cross-channel correlation. When compared to conventional ensemble averaging, the beamformer reduces the number of response repetitions required to achieve a given SNR by a factor which approaches the number of channels in the acquisition system.

Medical subject headings